It goes without saying that the results can be disastrous if calibration is conducted incorrectly or if the gas is not appropriate for the instrument being tested. By exposing the detector to a known concentration calibration gas, you can confirm the accuracy of the reading and the proper functioning of the alarm.Įmployers that fail to regularly test and document instrument performance are vulnerable to regulatory fines and the occurrence of serious and preventable workplace accidents.Ĭalibration Is Only as Accurate as the Test Gas If ever in doubt, conduct a full calibration. sensor poisons, high gas concentrations, extreme temperature, mechanical shock or stress) are suspected to be present Calibration should be conducted more frequently if environmental conditions could affect device performance (e.g.A full calibration must be carried out on any instrument that fails the functional test.A functional test (also known as a bump test) or full calibration should be performed each day before using direct reading portable gas monitor.